术前角膜散光轴向对人工晶状体植入术后等效球镜屈光度的影响

原文标题: Investigating the impact of preoperative corneal astigmatism orientation on the postoperative spherical equivalent refraction following intraocular lens implantation

Richard N. McNeely, Salissou Moutari, Emmanuel Eric Pazo, Jonathan Edward Moore

Eye and Vision, 2018 · 期刊论文 · doi:10.1186/s40662-018-0103-4 · 被引 3 次 · 开放获取(CC BY)

眼表与干眼白内障与眼前节手术

图文摘要. 回顾性研究 · 339 只眼. SE 预测误差(逆规散光,D): -0.132; SE 预测误差(顺规散光,D): 0.026. 角膜散光轴向会轻微改变人工晶状体的预测误差,但未达到有临床意义的程度。
图文摘要: SVG · PNG · JSON · 可在注明出处的前提下自由使用(CC BY 4.0)

要点速览

角膜散光轴向会轻微改变人工晶状体的预测误差,但未达到有临床意义的程度。

研究设计
回顾性研究
样本
339 只眼
SE 预测误差(逆规散光,D)
-0.132
SE 预测误差(顺规散光,D)
0.026

摘要

背景:探讨术前角膜散光轴向对单焦点人工晶状体(IOL)植入术后实现目标屈光度的影响。方法:本研究纳入 339 只接受顺利的白内障手术或屈光性晶状体置换术(RLE)并随后植入单焦点 IOL 的眼。首先按眼轴长度对眼进行分类,再按术前角膜前表面散光的轴向分类:第 1 组为逆规(ATR)角膜前表面散光,第 2 组为顺规(WTR)角膜前表面散光,第 3 组为斜轴(OB)角膜前表面散光。术前角膜散光由 IOLMaster(Carl Zeiss Meditec AG)测定。所有眼均完成术后验光,并分别计算和比较各组结果。结果:在眼轴长度大于 22.0 mm 且小于 25.0 mm 的眼中,第 2 组与第 3 组术前角膜散光的大小存在显著差异,第 2 组为 0.827 ± 0.376 D,第 3 组为 0.677 ± 0.387 D。术后等效球镜(SE)平均预测误差在第 1 组为 - 0.132 ± 0.475 D,第 2 组为 0.026 ± 0.497 D,第 3 组为 - 0.130 ± 0.477 D。第 1 组与第 2 组之间存在显著差异。在眼轴长度小于或等于 22.0 mm 以及大于或等于 25.0 mm 的眼中,各角膜前表面散光轴向组之间的术前角膜散光大小和术后 SE 预测误差均无显著差异。结论:在眼轴长度介于 22.0 mm 与 25.0 mm 之间、散光为 1.75 D 或以下的眼中,术前角膜前表面散光轴向显著影响术后生物测量预测误差,但该结果无临床意义。

译自英文原文摘要(许可协议:CC BY)。

显示英文原文摘要

BACKGROUND: To investigate the impact of the orientation of preoperative corneal astigmatism on achieving the postoperative target refraction following monofocal intraocular lens (IOL) implantation. METHODS: This study enrolled 339 eyes who had uneventful cataract surgery or refractive lens exchange (RLE) with subsequent monofocal IOL implantation. Eyes were initially categorized dependent upon axial length and then on the orientation of preoperative anterior corneal astigmatism. Group 1 had against-the-rule (ATR) anterior corneal astigmatism, group 2 had with-the-rule (WTR) anterior corneal astigmatism, and group 3 had oblique (OB) anterior corneal astigmatism. The preoperative corneal astigmatism was determined by the IOLMaster (Carl Zeiss Meditec AG). Postoperative refraction was completed for all eyes, and the results were calculated and compared for the separate groups. RESULTS: In eyes with axial lengths greater than 22.0 mm and less than 25.0 mm there was a significant difference between the magnitude of preoperative corneal astigmatism between groups 2 and 3 with 0.827 ± 0.376 D in group 2, and 0.677 ± 0.387 D in group 3. The mean postoperative spherical equivalent (SE) prediction error was - 0.132 ± 0.475 D in group 1, 0.026 ± 0.497 D in group 2, and - 0.130 ± 0.477 D in group 3. There was a significant difference between groups 1 and 2. There was no significant difference in the magnitude of preoperative corneal astigmatism and postoperative SE prediction error between the anterior corneal astigmatism orientation groups in eyes with axial lengths of less than or equal to 22.0 mm and greater than or equal to 25.0 mm. CONCLUSIONS: The orientation of preoperative anterior corneal astigmatism significantly affected the postoperative biometry prediction error in eyes with astigmatism of 1.75 D or less in eyes with the axial length between 22.0 mm and 25.0 mm. However, the results were not clinically significant.

关键词: monofocal intraocular lens; refractive lens exchange; cataract surgery; spherical equivalent refraction; axial length; against-the-rule astigmatism; with-the-rule astigmatism; oblique astigmatism

获取全文

引用格式

McNeely, R. N., Moutari, S., Pazo, E. E., Moore, J. E. (2018). Investigating the impact of preoperative corneal astigmatism orientation on the postoperative spherical equivalent refraction following intraocular lens implantation. Eye and Vision. https://doi.org/10.1186/s40662-018-0103-4

显示 BibTeX
@article{mcneely2018investigating,
  title = {{Investigating the impact of preoperative corneal astigmatism orientation on the postoperative spherical equivalent refraction following intraocular lens implantation}},
  author = {McNeely, Richard N. and Moutari, Salissou and Pazo, Emmanuel Eric and Moore, Jonathan Edward},
  journal = {Eye and Vision},
  year = {2018},
  doi = {10.1186/s40662-018-0103-4},
  url = {https://doi.org/10.1186/s40662-018-0103-4},
}

相关研究