人工泪液对干眼与非干眼患者眼部生物学参数影响的评估
原文标题: To evaluate the effects of artificial tears on ocular biological parameters in dry eye and non-dry eye patients
Yimeng Chen, Mingze Li, Jiayan Chen, Jiahui Zhao, Emmanuel Eric Pazo, Guanghao Qin, Xingru He
要点速览
人工泪液会改变眼轴长度和角膜厚度的测量值;滴用后应至少等待 5 分钟再进行生物测量。
- 研究设计
- 前瞻性研究
- 样本
- 98 只眼
- 滴药后建议等待时间
- 5 min
- 所有时间点的中央角膜厚度变化
- p < 0.001
摘要
本文摘要以原文(英文)显示,因为该文章的许可协议(CC BY-NC-ND)不允许发布译文。
To analyze the changes in ocular biometry readings in patients with dry eye (DE) and non-dry eye (NDE) after instilling artificial tear drop (ATD). A prospective study was conducted on 98 eyes of 98 patients (DE group, n = 54 eyes; NDE group, n = 44 eyes) who underwent ocular biometry with four consecutive measurements using the IOL Master 700. The baseline flat meridian of the anterior corneal surface (K1), steep meridian of the anterior corneal surface (K2), corneal curvature radius(R1, R2), axial length (AL), anterior chamber depth (ACD), central corneal thickness (CCT), white-to-white corneal diameter (WTW), pupil diameter, non-invasive tear break-up time (NITBUT) and tear meniscus height, (TMH) was measured at baseline, 30-s, 2-min, and 5-min intervals after instilling one drop of ATD. Variability in biometry measurements was assessed and compared. The mean age of the NDE group and DE group was 27.09 ± 10.21 years (30 females, 68.18%) and 29.96 ± 12.08 years (38 females, 70.37%), respectively. Following the application of ATD, intergroup comparisons revealed significant changes in K1 at 2 min and 5 min (p = 0.043, p = 0.043, and p = 0.038, respectively), and significant changes in R1 and R2 at 2 min. Intragroup comparisons showed that AL exhibited significant changes at 30 s (p < 0.001), while CCT demonstrated significant alterations across all time points (p < 0.001). Changes in ACD were observed at 30 s and 2 min in the dry eye group, and pupil diameter showed significant changes at 2 min in both the DE and NDE groups. No significant changes were detected in WTW at any point (all p > 0.05). Artificial tears significantly affect the readings of AL and CCT during ocular biometry. It is recommended to wait for at least 5 min after the installation of eye drops before performing ocular biometry measurements.
关键词: artificial tears; dry eye; ocular biometry; axial length; central corneal thickness; anterior chamber depth; corneal curvature; pupil size
获取全文
引用格式
显示 BibTeX
@article{chen2025to,
title = {{To evaluate the effects of artificial tears on ocular biological parameters in dry eye and non-dry eye patients}},
author = {Chen, Yimeng and Li, Mingze and Chen, Jiayan and Zhao, Jiahui and Pazo, Emmanuel Eric and Qin, Guanghao and He, Xingru},
journal = {Scientific Reports},
year = {2025},
doi = {10.1038/s41598-025-95801-5},
url = {https://doi.org/10.1038/s41598-025-95801-5},
}
相关研究
- 飞秒激光辅助 LASIK 术后角膜瓣边缘间隙对角膜神经再生影响的分析 (2026)
- 强脉冲光与睑缘清洁对泪膜和眼表影响的比较:一项随机对照研究 (2026)
- Effects of 3% diquafosol sodium on tear film stability, corneal aberration parameters, and vision-related quality of life in pediatric orthokeratology-associated dry eye disease (2026)
- 母亲孕前体重指数与子代散光及角膜曲率的关联:一项横断面研究 (2025)
- 过敏性结膜炎相关干眼中泪液淋巴毒素-α、免疫球蛋白 E 及基质金属蛋白酶-9 的差异表达 (2025)