ドライアイ患者と非ドライアイ患者における人工涙液の眼生体パラメータへの影響の評価
原題: To evaluate the effects of artificial tears on ocular biological parameters in dry eye and non-dry eye patients
Yimeng Chen, Mingze Li, Jiayan Chen, Jiahui Zhao, Emmanuel Eric Pazo, Guanghao Qin, Xingru He
要点
人工涙液は眼軸長と角膜厚の測定値を変化させるため、生体計測の前に少なくとも 5 分待つべきである。
- 研究デザイン
- 前向き研究
- 対象
- 98 眼
- 点眼後の推奨待機時間
- 5 min
- 全時点での CCT の変化
- p < 0.001
抄録
この抄録は原語(英語)で表示しています。論文のライセンス(CC BY-NC-ND)では翻訳の公開が認められていないためです。
To analyze the changes in ocular biometry readings in patients with dry eye (DE) and non-dry eye (NDE) after instilling artificial tear drop (ATD). A prospective study was conducted on 98 eyes of 98 patients (DE group, n = 54 eyes; NDE group, n = 44 eyes) who underwent ocular biometry with four consecutive measurements using the IOL Master 700. The baseline flat meridian of the anterior corneal surface (K1), steep meridian of the anterior corneal surface (K2), corneal curvature radius(R1, R2), axial length (AL), anterior chamber depth (ACD), central corneal thickness (CCT), white-to-white corneal diameter (WTW), pupil diameter, non-invasive tear break-up time (NITBUT) and tear meniscus height, (TMH) was measured at baseline, 30-s, 2-min, and 5-min intervals after instilling one drop of ATD. Variability in biometry measurements was assessed and compared. The mean age of the NDE group and DE group was 27.09 ± 10.21 years (30 females, 68.18%) and 29.96 ± 12.08 years (38 females, 70.37%), respectively. Following the application of ATD, intergroup comparisons revealed significant changes in K1 at 2 min and 5 min (p = 0.043, p = 0.043, and p = 0.038, respectively), and significant changes in R1 and R2 at 2 min. Intragroup comparisons showed that AL exhibited significant changes at 30 s (p < 0.001), while CCT demonstrated significant alterations across all time points (p < 0.001). Changes in ACD were observed at 30 s and 2 min in the dry eye group, and pupil diameter showed significant changes at 2 min in both the DE and NDE groups. No significant changes were detected in WTW at any point (all p > 0.05). Artificial tears significantly affect the readings of AL and CCT during ocular biometry. It is recommended to wait for at least 5 min after the installation of eye drops before performing ocular biometry measurements.
キーワード: artificial tears; dry eye; ocular biometry; axial length; central corneal thickness; anterior chamber depth; corneal curvature; pupil size
アクセス
引用方法
BibTeX を表示
@article{chen2025to,
title = {{To evaluate the effects of artificial tears on ocular biological parameters in dry eye and non-dry eye patients}},
author = {Chen, Yimeng and Li, Mingze and Chen, Jiayan and Zhao, Jiahui and Pazo, Emmanuel Eric and Qin, Guanghao and He, Xingru},
journal = {Scientific Reports},
year = {2025},
doi = {10.1038/s41598-025-95801-5},
url = {https://doi.org/10.1038/s41598-025-95801-5},
}
関連する論文
- フェムトセカンドレーザー併用 LASIK 後の角膜神経再生に対するフラップ縁間隙の影響の解析 (2026)
- 強パルス光と眼瞼清拭が涙液層と眼表面に及ぼす効果の比較:ランダム化比較試験 (2026)
- Effects of 3% diquafosol sodium on tear film stability, corneal aberration parameters, and vision-related quality of life in pediatric orthokeratology-associated dry eye disease (2026)
- 母親の妊娠前 BMI と子の乱視および角膜曲率との関連:横断研究 (2025)
- アレルギー性結膜炎に伴うドライアイにおける涙液中リンホトキシン α、免疫グロブリン E、マトリックスメタロプロテイナーゼ-9 の発現の違い (2025)