Validation of the Eyerobo FC Portable Fundus Camera for Diabetic Retinopathy Screening Using Public Datasets and Deep Learning
Emmanuel Eric Pazo, Xiangying Liu, Shoukuan Liu, Qinru Zhang, Mingxuan Zhang, Zhihui Zhang, Salissou Moutari, Muhammad Usama, Xinjun Ren
Retina & vascular imagingAI & deep learning
At a glance
A portable fundus camera with AI trained only on desktop images was noninferior to desktop systems for diabetic retinopathy screening.
- Design
- Prospective validation study
- Sample
- 104 eyes
- Sensitivity, portable camera
- 92.3%
- Specificity, portable camera
- 94.2%
- AUC (desktop benchmark 0.952)
- 0.977
See the results in 4 charts ↓ data as CSV · JSON
Abstract
To validate the diagnostic performance of the Eyerobo FC, a new portable non-mydriatic fundus camera for diabetic retinopathy (DR) screening, against an established desktop fundus camera benchmark using a transfer-learning approach in which artificial intelligence (AI)-based detection algorithms trained on desktop images were applied to Eyerobo FC images. This prospective validation study employed a three-tier experimental design. Tier 1 involved training a deep learning model (EfficientNet-B4) on standard desktop camera images from EyePACS and APTOS 2019 datasets. Tier 2 established the reference standard by evaluating the trained model on the Messidor-2 dataset (N = 1748 eyes) captured with a Topcon TRC NW6 desktop camera (sensitivity 92.7%, 95% CI 91.2–94.2%; AUC 0.952, 95% CI 0.943–0.961). Tier 3 validated the same AI model (without retraining) on images from the Eyerobo FC in a prospective cohort (N = 104 eyes: 52 referable DR, 52 non-referable). The primary outcome was noninferiority of sensitivity and specificity (margin $$\Delta$$ = 10%) compared with the desktop benchmark. Statistical analysis included bootstrap resampling (1000 iterations) for confidence intervals and a one-sided Z-test for the difference of proportions to assess noninferiority. In a balanced cohort of 104 eyes (52 referable DR, 52 non-referable), the Eyerobo FC achieved sensitivity of 92.3% (95% CI 84.4–98.2%) and specificity of 94.2% (95% CI 87.0–100%), demonstrating noninferior performance compared with the desktop benchmark (sensitivity 92.7%, specificity 94.3%). The sensitivity difference of $$-0.4$$ percentage points and the specificity difference of $$-0.1$$ percentage points were both within the noninferiority margin. AUC was 0.977 (95% CI 0.945–0.997) versus 0.952 for the desktop benchmark. The AI model correctly classified 97 of 104 eyes (93.3% accuracy, 95% CI 88.5–98.1%), with 4 false negatives and 3 false positives. Noninferiority was statistically confirmed for both sensitivity and specificity (P < 0.05). Inter-grader agreement was excellent (Cohen’s kappa = 0.917). Nonmydriatic image gradability rate was 94.4%. Grad-CAM visualization confirmed appropriate model attention to hemorrhages, exudates, and microaneurysms rather than artifacts. The Eyerobo fundus camera demonstrates noninferior diagnostic performance (sensitivity 92.3%, specificity 94.2%, AUC 0.977) compared with desktop systems when evaluated with AI algorithms trained exclusively on desktop images. These findings support deploying portable AI-assisted screening in resource-constrained and point-of-care settings, with successful cross-domain transfer learning enabling algorithmic generalizability across imaging platforms.
Keywords: diabetic retinopathy screening; transfer learning; deep learning; EfficientNet-B4; non-inferiority; referable diabetic retinopathy; Messidor-2; EyePACS
Results in charts
4 charts · 3-minute read
Diabetic retinopathy screening usually relies on desktop fundus cameras that need a fixed clinic and, often, dilated pupils. This study tested whether a portable non-mydriatic camera, the Eyerobo FC, can stand in for them when an AI model reads the images. An EfficientNet-B4 model trained only on desktop images (EyePACS and APTOS) was benchmarked on 1,748 Messidor-2 desktop images, then applied without retraining to Eyerobo FC images of 104 eyes. These four charts are built from the paper's published tables.
1. As accurate as the desktop camera
The same AI model, never shown a portable-camera image, read 104 eyes photographed with the Eyerobo FC. Sensitivity was 92.3% against 92.7% on the desktop benchmark, and specificity 94.2% against 94.3%; both differences fell within the pre-set 10-point noninferiority margin. The Eyerobo FC's intervals are much wider because its cohort is about 17 times smaller (104 against 1,748 eyes).
Data table
| Metric | Desktop (Messidor-2) (95% CI) | Eyerobo FC (95% CI) |
|---|---|---|
| Sensitivity | 0.927 [0.912, 0.942] | 0.923 [0.844, 0.982] |
| Specificity | 0.943 [0.931, 0.955] | 0.942 [0.870, 1.000] |
| Accuracy | 0.935 [0.924, 0.946] | 0.933 [0.885, 0.971] |
| F1-score | 0.931 [0.919, 0.943] | 0.932 [0.873, 0.975] |
| AUC | 0.952 [0.943, 0.961] | 0.977 [0.945, 0.997] |
2. Consistent across severity and image quality
Sensitivity was 94.4% for moderate non-proliferative retinopathy and 83.3% for both severe non-proliferative and proliferative disease, and 93.9% on excellent-quality images against 85.7% on good or adequate ones. With 6 to 36 eyes per subgroup the intervals are wide, and the paper reports no significant differences between subgroups (all P > 0.05).
Data table
| Subgroup | Sensitivity (95% CI) |
|---|---|
| All referable DR | 88.9 [77.8, 97.4] |
| Moderate NPDR | 94.4 [83.9, 100.0] |
| Severe NPDR | 83.3 [62.6, 97.7] |
| Proliferative DR | 83.3 [52.5, 98.0] |
| DR only | 90.9 [80.0, 98.3] |
| DR + DME | 85.7 [69.7, 96.5] |
| Excellent quality | 93.9 [85.2, 98.9] |
| Good/adequate | 85.7 [73.8, 94.9] |
3. Slightly lower image quality without dilation
Without pupil dilation, 94.4% of Eyerobo FC images were gradable, against 98.4% for the dilated desktop camera (P = 0.082). Portable images were less often rated excellent (45.8% against 62.3%, P = 0.031), less often free of artifacts (76.4% against 87.2%, P = 0.048; mostly eyelash shadows and peripheral reflections) and less often showed vessel detail clearly (87.5% against 95.6%, P = 0.029).
Data table
| Quality metric | Eyerobo FC (no dilation) | Topcon (dilated) |
|---|---|---|
| Gradable, overall | 94.4 | 98.4 |
| Excellent (score 3) | 45.8 | 62.3 |
| Both fields captured | 95.8 | 98.9 |
| No artifacts | 76.4 | 87.2 |
| Clear disc margins | 91.7 | 97.8 |
| Clear vessel detail | 87.5 | 95.6 |
| Clear macula | 93.1 | 98.3 |
4. Graders and AI agree
Two retina specialists agreed on referable disease in 69 of 72 eyes (kappa 0.917), and each agreed with the AI model at kappa 0.833 and 0.806. For individual lesions, kappa ranged from 0.833 for hemorrhages to 1.000 for vitreous hemorrhage. Every point estimate is at least 0.80, inside the paper's 'excellent' band, though several lower confidence bounds fall below 0.75.
Data table
| Comparison | Cohen's kappa (95% CI) |
|---|---|
| Grader 1 vs grader 2 (referable DR) | 0.917 [0.841, 0.982] |
| Grader 1 vs AI model (referable DR) | 0.833 [0.731, 0.924] |
| Grader 2 vs AI model (referable DR) | 0.806 [0.698, 0.905] |
| Grader 1 vs grader 2 (5-level severity) | 0.842 [0.752, 0.926] |
| Microaneurysms | 0.862 [0.773, 0.941] |
| Hemorrhages | 0.833 [0.736, 0.924] |
| Hard exudates | 0.917 [0.845, 0.981] |
| Cotton-wool spots | 0.943 [0.881, 0.991] |
| Neovascularization | 0.972 [0.927, 0.998] |
| Vitreous hemorrhage | 1.000 [1.000, 1.000] |
| Overall gradability | 0.831 [0.686, 0.951] |
| Quality score (4-point) | 0.822 [0.726, 0.911] |
What it means
With an AI model trained only on desktop images, the portable Eyerobo FC matched the desktop camera's sensitivity and specificity for referable diabetic retinopathy, despite somewhat lower image quality without dilation.
Limitation: The Eyerobo FC cohort is small (104 eyes) and enriched to 50% referable disease, well above typical screening populations (10–30%), which affects predictive values; subgroups hold 6 to 36 eyes.
Data and reuse
All values are copied from Tables 1–5 of the published article (open access under CC BY-NC 4.0). Download every value as tidy CSV (one measurement per row) or JSON. Charts and data follow the article's licence (CC BY-NC 4.0): reuse with attribution, non-commercially. For full methods, read the full article.
All published results tables (5)
| Characteristic | Eyerobo FC cohort | Messidor-2 cohort |
|---|---|---|
| Total eyes | 104 | 1748 |
| Unique patients | 89 | 874 |
| Right eye (OD) | 58 (55.8%) | 874 (50.0%) |
| Left eye (OS) | 46 (44.2%) | 874 (50.0%) |
| Referable DR | 52 (50.0%) | 874 (50.0%) |
| Nonreferable | 52 (50.0%) | 874 (50.0%) |
| Camera model | Eyerobo FC iFLASH-I PRO | Topcon TRC NW6 |
| Imaging type | Portable, non-mydriatic | Desktop, mydriatic |
| Field of view | 45° | 45° |
| Image resolution | 2048 × 1536 pixels | 2240 × 1488 pixels |
| AI model | EfficientNet-B4 | EfficientNet-B4 |
| Training data | EyePACS + APTOS | EyePACS + APTOS |
| Reference standard | Human grader | Human grader |
| Metric | Desktop camera (tier 2, Messidor-2, 1748 eyes) | Desktop camera (tier 2, Messidor-2, 1748 eyes): 95% CI lower | Desktop camera (tier 2, Messidor-2, 1748 eyes): 95% CI upper | Eyerobo FC (tier 3, 104 eyes) | Eyerobo FC (tier 3, 104 eyes): 95% CI lower | Eyerobo FC (tier 3, 104 eyes): 95% CI upper |
|---|---|---|---|---|---|---|
| Sensitivity | 0.927 | 0.912 | 0.942 | 0.923 | 0.844 | 0.982 |
| Specificity | 0.943 | 0.931 | 0.955 | 0.942 | 0.87 | 1 |
| Accuracy | 0.935 | 0.924 | 0.946 | 0.933 | 0.885 | 0.971 |
| F1-score | 0.931 | 0.919 | 0.943 | 0.932 | 0.873 | 0.975 |
| AUC | 0.952 | 0.943 | 0.961 | 0.977 | 0.945 | 0.997 |
| Quality metric | Eyerobo FC (%) | Eyerobo FC n | Eyerobo FC total | Topcon TRC NW6 (%) | P-value |
|---|---|---|---|---|---|
| Gradable, overall | 94.4 | 68 | 72 | 98.4 | 0.082 |
| Gradable, with mydriasis | – | – | – | 98.4 | – |
| Gradable, without mydriasis | 94.4 | – | – | – | – |
| Excellent (score 3) | 45.8 | 33 | 72 | 62.3 | 0.031 |
| Good (score 2) | 48.6 | 35 | 72 | 36.1 | – |
| Adequate (score 1) | 5.6 | 4 | 72 | 1.6 | – |
| Ungradable (score 0) | 5.6 | 4 | 72 | 1.6 | – |
| Both fields captured | 95.8 | 69 | 72 | 98.9 | 0.157 |
| Single field only | 4.2 | 3 | 72 | 1.1 | – |
| No artifacts | 76.4 | 55 | 72 | 87.2 | 0.048 |
| Minor artifacts | 18.1 | 13 | 72 | 11.2 | – |
| Major artifacts | 5.6 | 4 | 72 | 1.6 | – |
| Clear disc margins | 91.7 | 66 | 72 | 97.8 | 0.064 |
| Clear vessel detail | 87.5 | 63 | 72 | 95.6 | 0.029 |
| Clear macula | 93.1 | 67 | 72 | 98.3 | 0.074 |
| Subgroup | N (eyes) | Sensitivity (%) | Sensitivity 95% CI lower | Sensitivity 95% CI upper | Specificity (%) | Specificity 95% CI lower | Specificity 95% CI upper | PPV (%) | PPV 95% CI lower | PPV 95% CI upper | NPV (%) | NPV 95% CI lower | NPV 95% CI upper |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| All referable DR | 36 | 88.9 | 77.8 | 97.4 | 94.4 | 86.5 | 100 | 94.1 | 84.8 | 99.3 | 89.5 | 79.3 | 96.1 |
| Moderate NPDR | 18 | 94.4 | 83.9 | 100 | 94.4 | 86.5 | 100 | 89.5 | 75.2 | 97.9 | 97.1 | 91.6 | 100 |
| Severe NPDR | 12 | 83.3 | 62.6 | 97.7 | 94.4 | 86.5 | 100 | 90.9 | 75.7 | 98.1 | 89.5 | 79.3 | 96.1 |
| Proliferative DR | 6 | 83.3 | 52.5 | 98 | 94.4 | 86.5 | 100 | 83.3 | 52.5 | 98 | 94.4 | 86.5 | 100 |
| DR only | 22 | 90.9 | 80 | 98.3 | 94.4 | 86.5 | 100 | 95.2 | 86.5 | 99 | 89.5 | 79.3 | 96.1 |
| DR + DME | 14 | 85.7 | 69.7 | 96.5 | 94.4 | 86.5 | 100 | 92.3 | 79.1 | 98.4 | 89.5 | 79.3 | 96.1 |
| Excellent quality | 33 | 93.9 | 85.2 | 98.9 | 97 | 91.5 | 100 | 96.9 | 89.3 | 99.6 | 94.1 | 86.5 | 98.4 |
| Good/adequate | 35 | 85.7 | 73.8 | 94.9 | 91.4 | 82.3 | 97.1 | 90.9 | 80 | 97.7 | 86.5 | 75.7 | 94.4 |
| Comparison | Agreement (%) | Agreeing eyes | Eyes | Cohen's kappa | Kappa 95% CI lower | Kappa 95% CI upper |
|---|---|---|---|---|---|---|
| Grader 1 vs grader 2 (referable DR) | 95.8 | 69 | 72 | 0.917 | 0.841 | 0.982 |
| Grader 1 vs AI model (referable DR) | 91.7 | 66 | 72 | 0.833 | 0.731 | 0.924 |
| Grader 2 vs AI model (referable DR) | 90.3 | 65 | 72 | 0.806 | 0.698 | 0.905 |
| Grader 1 vs grader 2 (5-level severity) | 88.9 | 64 | 72 | 0.842 | 0.752 | 0.926 |
| Microaneurysms | 93.1 | 67 | 72 | 0.862 | 0.773 | 0.941 |
| Hemorrhages | 91.7 | 66 | 72 | 0.833 | 0.736 | 0.924 |
| Hard exudates | 95.8 | 69 | 72 | 0.917 | 0.845 | 0.981 |
| Cotton-wool spots | 97.2 | 70 | 72 | 0.943 | 0.881 | 0.991 |
| Neovascularization | 98.6 | 71 | 72 | 0.972 | 0.927 | 0.998 |
| Vitreous hemorrhage | 100 | 72 | 72 | 1 | 1 | 1 |
| Overall gradability | 97.2 | 70 | 72 | 0.831 | 0.686 | 0.951 |
| Quality score (4-point) | 88.9 | 64 | 72 | 0.822 | 0.726 | 0.911 |
Access
How to cite
Show BibTeX
@article{pazo2026validation,
title = {{Validation of the Eyerobo FC Portable Fundus Camera for Diabetic Retinopathy Screening Using Public Datasets and Deep Learning}},
author = {Pazo, Emmanuel Eric and Liu, Xiangying and Liu, Shoukuan and Zhang, Qinru and Zhang, Mingxuan and Zhang, Zhihui and Moutari, Salissou and Usama, Muhammad and Ren, Xinjun},
journal = {Ophthalmology and Therapy},
year = {2026},
doi = {10.1007/s40123-026-01353-w},
url = {https://doi.org/10.1007/s40123-026-01353-w},
}
Related work
- Association of Anti-VEGF Knowledge and Attitudes With Self-Reported Treatment Interruption: A Temporally Separated Classification Study (2026)
- Automated deep learning-based retinoschisis and detachment volume measurement in pathological myopia with posterior scleral contraction (2026)
- Development of a predictive model based on aqueous cytokines data for response to anti-VEGF therapy in diabetic macular edema (2026)
- Structure-preserving super-resolution of retinal fundus images via a dual-transformer residual network (2026)
- Exploring the effect of hypertension on retinal microvasculature using deep learning on East Asian population (2020)