Validation of the Eyerobo FC Portable Fundus Camera for Diabetic Retinopathy Screening Using Public Datasets and Deep Learning

Emmanuel Eric Pazo, Xiangying Liu, Shoukuan Liu, Qinru Zhang, Mingxuan Zhang, Zhihui Zhang, Salissou Moutari, Muhammad Usama, Xinjun Ren

Ophthalmology and Therapy, 2026 · article · doi:10.1007/s40123-026-01353-w · cited by 3 · Open access (CC BY-NC)

Retina & vascular imagingAI & deep learning

Visual abstract. Prospective validation study of 104 eyes. Key results: Sensitivity, portable camera: 92.3%; Specificity, portable camera: 94.2%; AUC (desktop benchmark 0.952): 0.977. Conclusion: A portable fundus camera with AI trained only on desktop images was noninferior to desktop systems for diabetic retinopathy screening.
Visual abstract: SVG · PNG · JSON · free to reuse with attribution (CC BY 4.0)

At a glance

A portable fundus camera with AI trained only on desktop images was noninferior to desktop systems for diabetic retinopathy screening.

Design
Prospective validation study
Sample
104 eyes
Sensitivity, portable camera
92.3%
Specificity, portable camera
94.2%
AUC (desktop benchmark 0.952)
0.977

See the results in 4 charts ↓ data as CSV · JSON

Abstract

To validate the diagnostic performance of the Eyerobo FC, a new portable non-mydriatic fundus camera for diabetic retinopathy (DR) screening, against an established desktop fundus camera benchmark using a transfer-learning approach in which artificial intelligence (AI)-based detection algorithms trained on desktop images were applied to Eyerobo FC images. This prospective validation study employed a three-tier experimental design. Tier 1 involved training a deep learning model (EfficientNet-B4) on standard desktop camera images from EyePACS and APTOS 2019 datasets. Tier 2 established the reference standard by evaluating the trained model on the Messidor-2 dataset (N = 1748 eyes) captured with a Topcon TRC NW6 desktop camera (sensitivity 92.7%, 95% CI 91.2–94.2%; AUC 0.952, 95% CI 0.943–0.961). Tier 3 validated the same AI model (without retraining) on images from the Eyerobo FC in a prospective cohort (N = 104 eyes: 52 referable DR, 52 non-referable). The primary outcome was noninferiority of sensitivity and specificity (margin $$\Delta$$ = 10%) compared with the desktop benchmark. Statistical analysis included bootstrap resampling (1000 iterations) for confidence intervals and a one-sided Z-test for the difference of proportions to assess noninferiority. In a balanced cohort of 104 eyes (52 referable DR, 52 non-referable), the Eyerobo FC achieved sensitivity of 92.3% (95% CI 84.4–98.2%) and specificity of 94.2% (95% CI 87.0–100%), demonstrating noninferior performance compared with the desktop benchmark (sensitivity 92.7%, specificity 94.3%). The sensitivity difference of $$-0.4$$ percentage points and the specificity difference of $$-0.1$$ percentage points were both within the noninferiority margin. AUC was 0.977 (95% CI 0.945–0.997) versus 0.952 for the desktop benchmark. The AI model correctly classified 97 of 104 eyes (93.3% accuracy, 95% CI 88.5–98.1%), with 4 false negatives and 3 false positives. Noninferiority was statistically confirmed for both sensitivity and specificity (P < 0.05). Inter-grader agreement was excellent (Cohen’s kappa = 0.917). Nonmydriatic image gradability rate was 94.4%. Grad-CAM visualization confirmed appropriate model attention to hemorrhages, exudates, and microaneurysms rather than artifacts. The Eyerobo fundus camera demonstrates noninferior diagnostic performance (sensitivity 92.3%, specificity 94.2%, AUC 0.977) compared with desktop systems when evaluated with AI algorithms trained exclusively on desktop images. These findings support deploying portable AI-assisted screening in resource-constrained and point-of-care settings, with successful cross-domain transfer learning enabling algorithmic generalizability across imaging platforms.

Keywords: diabetic retinopathy screening; transfer learning; deep learning; EfficientNet-B4; non-inferiority; referable diabetic retinopathy; Messidor-2; EyePACS

Results in charts

4 charts · 3-minute read

Diabetic retinopathy screening usually relies on desktop fundus cameras that need a fixed clinic and, often, dilated pupils. This study tested whether a portable non-mydriatic camera, the Eyerobo FC, can stand in for them when an AI model reads the images. An EfficientNet-B4 model trained only on desktop images (EyePACS and APTOS) was benchmarked on 1,748 Messidor-2 desktop images, then applied without retraining to Eyerobo FC images of 104 eyes. These four charts are built from the paper's published tables.

1. As accurate as the desktop camera

The same AI model, never shown a portable-camera image, read 104 eyes photographed with the Eyerobo FC. Sensitivity was 92.3% against 92.7% on the desktop benchmark, and specificity 94.2% against 94.3%; both differences fell within the pre-set 10-point noninferiority margin. The Eyerobo FC's intervals are much wider because its cohort is about 17 times smaller (104 against 1,748 eyes).

Diagnostic performance: desktop camera against the portable Eyerobo FC Diagnostic performance: desktop camera against the portable Eyerobo FC (Referable diabetic retinopathy · point estimate with bootstrap 95% CI · higher is better). Sensitivity, Desktop (Messidor-2) 0.927 [0.912, 0.942]; Sensitivity, Eyerobo FC 0.923 [0.844, 0.982]; Specificity, Desktop (Messidor-2) 0.943 [0.931, 0.955]; Specificity, Eyerobo FC 0.942 [0.870, 1.000]; Accuracy, Desktop (Messidor-2) 0.935 [0.924, 0.946]; Accuracy, Eyerobo FC 0.933 [0.885, 0.971]; F1-score, Desktop (Messidor-2) 0.931 [0.919, 0.943]; F1-score, Eyerobo FC 0.932 [0.873, 0.975]; AUC, Desktop (Messidor-2) 0.952 [0.943, 0.961]; AUC, Eyerobo FC 0.977 [0.945, 0.997]. 0.80 0.85 0.90 0.95 1.00 Sensitivity Sensitivity · Desktop (Messidor-2): 0.927 [0.912, 0.942]Desktop (Messidor-2) Sensitivity · Eyerobo FC: 0.923 [0.844, 0.982]Eyerobo FC0.923 [0.844, 0.982] Specificity Specificity · Desktop (Messidor-2): 0.943 [0.931, 0.955]Desktop (Messidor-2) Specificity · Eyerobo FC: 0.942 [0.870, 1.000]Eyerobo FC0.942 [0.870, 1.000] Accuracy Accuracy · Desktop (Messidor-2): 0.935 [0.924, 0.946]Desktop (Messidor-2) Accuracy · Eyerobo FC: 0.933 [0.885, 0.971]Eyerobo FC0.933 [0.885, 0.971] F1-score F1-score · Desktop (Messidor-2): 0.931 [0.919, 0.943]Desktop (Messidor-2) F1-score · Eyerobo FC: 0.932 [0.873, 0.975]Eyerobo FC0.932 [0.873, 0.975] AUC AUC · Desktop (Messidor-2): 0.952 [0.943, 0.961]Desktop (Messidor-2) AUC · Eyerobo FC: 0.977 [0.945, 0.997]Eyerobo FC0.977 [0.945, 0.997]
Diagnostic performance: desktop camera against the portable Eyerobo FC. Referable diabetic retinopathy · point estimate with bootstrap 95% CI · higher is better. Source: Pazo et al., Ophthalmology and Therapy 2026, Table 2 · doi:10.1007/s40123-026-01353-w. Download: PNG · SVG
Data table
Diagnostic performance: desktop camera against the portable Eyerobo FC
MetricDesktop (Messidor-2) (95% CI)Eyerobo FC (95% CI)
Sensitivity0.927 [0.912, 0.942]0.923 [0.844, 0.982]
Specificity0.943 [0.931, 0.955]0.942 [0.870, 1.000]
Accuracy0.935 [0.924, 0.946]0.933 [0.885, 0.971]
F1-score0.931 [0.919, 0.943]0.932 [0.873, 0.975]
AUC0.952 [0.943, 0.961]0.977 [0.945, 0.997]

2. Consistent across severity and image quality

Sensitivity was 94.4% for moderate non-proliferative retinopathy and 83.3% for both severe non-proliferative and proliferative disease, and 93.9% on excellent-quality images against 85.7% on good or adequate ones. With 6 to 36 eyes per subgroup the intervals are wide, and the paper reports no significant differences between subgroups (all P > 0.05).

Eyerobo FC sensitivity by subgroup, with 95% confidence intervals Eyerobo FC sensitivity by subgroup, with 95% confidence intervals (Percent · 6 to 36 eyes per subgroup · line marks the desktop benchmark). All referable DR 88.9 [77.8, 97.4]; Moderate NPDR 94.4 [83.9, 100.0]; Severe NPDR 83.3 [62.6, 97.7]; Proliferative DR 83.3 [52.5, 98.0]; DR only 90.9 [80.0, 98.3]; DR + DME 85.7 [69.7, 96.5]; Excellent quality 93.9 [85.2, 98.9]; Good/adequate 85.7 [73.8, 94.9]. 50 60 70 80 90 100 Desktop sensitivity (92.7%) All referable DR: 88.9 [77.8, 97.4]All referable DR88.9 [77.8, 97.4] Moderate NPDR: 94.4 [83.9, 100.0]Moderate NPDR Severe NPDR: 83.3 [62.6, 97.7]Severe NPDR Proliferative DR: 83.3 [52.5, 98.0]Proliferative DR DR only: 90.9 [80.0, 98.3]DR only DR + DME: 85.7 [69.7, 96.5]DR + DME Excellent quality: 93.9 [85.2, 98.9]Excellent quality Good/adequate: 85.7 [73.8, 94.9]Good/adequate
Eyerobo FC sensitivity by subgroup, with 95% confidence intervals. Percent · 6 to 36 eyes per subgroup · line marks the desktop benchmark. Source: Pazo et al., Ophthalmology and Therapy 2026, Table 4 · doi:10.1007/s40123-026-01353-w. Download: PNG · SVG
Data table
Eyerobo FC sensitivity by subgroup, with 95% confidence intervals
SubgroupSensitivity (95% CI)
All referable DR88.9 [77.8, 97.4]
Moderate NPDR94.4 [83.9, 100.0]
Severe NPDR83.3 [62.6, 97.7]
Proliferative DR83.3 [52.5, 98.0]
DR only90.9 [80.0, 98.3]
DR + DME85.7 [69.7, 96.5]
Excellent quality93.9 [85.2, 98.9]
Good/adequate85.7 [73.8, 94.9]

3. Slightly lower image quality without dilation

Without pupil dilation, 94.4% of Eyerobo FC images were gradable, against 98.4% for the dilated desktop camera (P = 0.082). Portable images were less often rated excellent (45.8% against 62.3%, P = 0.031), less often free of artifacts (76.4% against 87.2%, P = 0.048; mostly eyelash shadows and peripheral reflections) and less often showed vessel detail clearly (87.5% against 95.6%, P = 0.029).

Image quality: Eyerobo FC without dilation against Topcon desktop with dilation Image quality: Eyerobo FC without dilation against Topcon desktop with dilation (Percent of images · Eyerobo FC 72 eyes · higher is better). Gradable, overall, Eyerobo FC (no dilation) 94.4; Gradable, overall, Topcon (dilated) 98.4; Excellent (score 3), Eyerobo FC (no dilation) 45.8; Excellent (score 3), Topcon (dilated) 62.3; Both fields captured, Eyerobo FC (no dilation) 95.8; Both fields captured, Topcon (dilated) 98.9; No artifacts, Eyerobo FC (no dilation) 76.4; No artifacts, Topcon (dilated) 87.2; Clear disc margins, Eyerobo FC (no dilation) 91.7; Clear disc margins, Topcon (dilated) 97.8; Clear vessel detail, Eyerobo FC (no dilation) 87.5; Clear vessel detail, Topcon (dilated) 95.6; Clear macula, Eyerobo FC (no dilation) 93.1; Clear macula, Topcon (dilated) 98.3. 40 50 60 70 80 90 100 Gradable, overall Gradable, overall · Eyerobo FC (no dilation): 94.4Eyerobo FC (no dilation)94.4 Gradable, overall · Topcon (dilated): 98.4Topcon (dilated) Excellent (score 3) Excellent (score 3) · Eyerobo FC (no dilation): 45.8Eyerobo FC (no dilation)45.8 Excellent (score 3) · Topcon (dilated): 62.3Topcon (dilated) Both fields captured Both fields captured · Eyerobo FC (no dilation): 95.8Eyerobo FC (no dilation)95.8 Both fields captured · Topcon (dilated): 98.9Topcon (dilated) No artifacts No artifacts · Eyerobo FC (no dilation): 76.4Eyerobo FC (no dilation)76.4 No artifacts · Topcon (dilated): 87.2Topcon (dilated) Clear disc margins Clear disc margins · Eyerobo FC (no dilation): 91.7Eyerobo FC (no dilation)91.7 Clear disc margins · Topcon (dilated): 97.8Topcon (dilated) Clear vessel detail Clear vessel detail · Eyerobo FC (no dilation): 87.5Eyerobo FC (no dilation)87.5 Clear vessel detail · Topcon (dilated): 95.6Topcon (dilated) Clear macula Clear macula · Eyerobo FC (no dilation): 93.1Eyerobo FC (no dilation)93.1 Clear macula · Topcon (dilated): 98.3Topcon (dilated)
Image quality: Eyerobo FC without dilation against Topcon desktop with dilation. Percent of images · Eyerobo FC 72 eyes · higher is better. Source: Pazo et al., Ophthalmology and Therapy 2026, Table 3 · doi:10.1007/s40123-026-01353-w. Download: PNG · SVG
Data table
Image quality: Eyerobo FC without dilation against Topcon desktop with dilation
Quality metricEyerobo FC (no dilation)Topcon (dilated)
Gradable, overall94.498.4
Excellent (score 3)45.862.3
Both fields captured95.898.9
No artifacts76.487.2
Clear disc margins91.797.8
Clear vessel detail87.595.6
Clear macula93.198.3

4. Graders and AI agree

Two retina specialists agreed on referable disease in 69 of 72 eyes (kappa 0.917), and each agreed with the AI model at kappa 0.833 and 0.806. For individual lesions, kappa ranged from 0.833 for hemorrhages to 1.000 for vitreous hemorrhage. Every point estimate is at least 0.80, inside the paper's 'excellent' band, though several lower confidence bounds fall below 0.75.

Agreement between expert graders and the AI model (Cohen's kappa) Agreement between expert graders and the AI model (Cohen's kappa) (72 eyes · bootstrap 95% CI · line marks the 0.75 threshold for excellent agreement). Grader 1 vs grader 2 (referable DR) 0.917 [0.841, 0.982]; Grader 1 vs AI model (referable DR) 0.833 [0.731, 0.924]; Grader 2 vs AI model (referable DR) 0.806 [0.698, 0.905]; Grader 1 vs grader 2 (5-level severity) 0.842 [0.752, 0.926]; Microaneurysms 0.862 [0.773, 0.941]; Hemorrhages 0.833 [0.736, 0.924]; Hard exudates 0.917 [0.845, 0.981]; Cotton-wool spots 0.943 [0.881, 0.991]; Neovascularization 0.972 [0.927, 0.998]; Vitreous hemorrhage 1.000 [1.000, 1.000]; Overall gradability 0.831 [0.686, 0.951]; Quality score (4-point) 0.822 [0.726, 0.911]. 0.6 0.7 0.8 0.9 1.0 Excellent (0.75) Grader 1 vs grader 2 (referable DR): 0.917 [0.841, 0.982]Grader 1 vs grader 2 (referable DR)0.917 [0.841, 0.982] Grader 1 vs AI model (referable DR): 0.833 [0.731, 0.924]Grader 1 vs AI model (referable DR)0.833 [0.731, 0.924] Grader 2 vs AI model (referable DR): 0.806 [0.698, 0.905]Grader 2 vs AI model (referable DR)0.806 [0.698, 0.905] Grader 1 vs grader 2 (5-level severity): 0.842 [0.752, 0.926]Grader 1 vs grader 2 (5-level severity)0.842 [0.752, 0.926] Microaneurysms: 0.862 [0.773, 0.941]Microaneurysms0.862 [0.773, 0.941] Hemorrhages: 0.833 [0.736, 0.924]Hemorrhages0.833 [0.736, 0.924] Hard exudates: 0.917 [0.845, 0.981]Hard exudates0.917 [0.845, 0.981] Cotton-wool spots: 0.943 [0.881, 0.991]Cotton-wool spots0.943 [0.881, 0.991] Neovascularization: 0.972 [0.927, 0.998]Neovascularization0.972 [0.927, 0.998] Vitreous hemorrhage: 1.000 [1.000, 1.000]Vitreous hemorrhage1.000 [1.000, 1.000] Overall gradability: 0.831 [0.686, 0.951]Overall gradability0.831 [0.686, 0.951] Quality score (4-point): 0.822 [0.726, 0.911]Quality score (4-point)0.822 [0.726, 0.911]
Agreement between expert graders and the AI model (Cohen's kappa). 72 eyes · bootstrap 95% CI · line marks the 0.75 threshold for excellent agreement. Source: Pazo et al., Ophthalmology and Therapy 2026, Table 5 · doi:10.1007/s40123-026-01353-w. Download: PNG · SVG
Data table
Agreement between expert graders and the AI model (Cohen's kappa)
ComparisonCohen's kappa (95% CI)
Grader 1 vs grader 2 (referable DR)0.917 [0.841, 0.982]
Grader 1 vs AI model (referable DR)0.833 [0.731, 0.924]
Grader 2 vs AI model (referable DR)0.806 [0.698, 0.905]
Grader 1 vs grader 2 (5-level severity)0.842 [0.752, 0.926]
Microaneurysms0.862 [0.773, 0.941]
Hemorrhages0.833 [0.736, 0.924]
Hard exudates0.917 [0.845, 0.981]
Cotton-wool spots0.943 [0.881, 0.991]
Neovascularization0.972 [0.927, 0.998]
Vitreous hemorrhage1.000 [1.000, 1.000]
Overall gradability0.831 [0.686, 0.951]
Quality score (4-point)0.822 [0.726, 0.911]

What it means

With an AI model trained only on desktop images, the portable Eyerobo FC matched the desktop camera's sensitivity and specificity for referable diabetic retinopathy, despite somewhat lower image quality without dilation.

Limitation: The Eyerobo FC cohort is small (104 eyes) and enriched to 50% referable disease, well above typical screening populations (10–30%), which affects predictive values; subgroups hold 6 to 36 eyes.

Data and reuse

All values are copied from Tables 1–5 of the published article (open access under CC BY-NC 4.0). Download every value as tidy CSV (one measurement per row) or JSON. Charts and data follow the article's licence (CC BY-NC 4.0): reuse with attribution, non-commercially. For full methods, read the full article.

All published results tables (5)
Table 1. Baseline characteristics of study cohorts.
CharacteristicEyerobo FC cohortMessidor-2 cohort
Total eyes1041748
Unique patients89874
Right eye (OD)58 (55.8%)874 (50.0%)
Left eye (OS)46 (44.2%)874 (50.0%)
Referable DR52 (50.0%)874 (50.0%)
Nonreferable52 (50.0%)874 (50.0%)
Camera modelEyerobo FC iFLASH-I PROTopcon TRC NW6
Imaging typePortable, non-mydriaticDesktop, mydriatic
Field of view45°45°
Image resolution2048 × 1536 pixels2240 × 1488 pixels
AI modelEfficientNet-B4EfficientNet-B4
Training dataEyePACS + APTOSEyePACS + APTOS
Reference standardHuman graderHuman grader
Table 2. Diagnostic performance comparison: Eyerobo FC (104 eyes) versus desktop camera (Messidor-2, 1748 eyes); point estimate and bootstrap 95% CI.
MetricDesktop camera (tier 2, Messidor-2, 1748 eyes)Desktop camera (tier 2, Messidor-2, 1748 eyes): 95% CI lowerDesktop camera (tier 2, Messidor-2, 1748 eyes): 95% CI upperEyerobo FC (tier 3, 104 eyes)Eyerobo FC (tier 3, 104 eyes): 95% CI lowerEyerobo FC (tier 3, 104 eyes): 95% CI upper
Sensitivity0.9270.9120.9420.9230.8440.982
Specificity0.9430.9310.9550.9420.871
Accuracy0.9350.9240.9460.9330.8850.971
F1-score0.9310.9190.9430.9320.8730.975
AUC0.9520.9430.9610.9770.9450.997
Table 3. Image quality assessment: Eyerobo FC (non-mydriatic, 72 eyes) versus Topcon TRC NW6 (mydriatic); P from Fisher's exact test.
Quality metricEyerobo FC (%)Eyerobo FC nEyerobo FC totalTopcon TRC NW6 (%)P-value
Gradable, overall94.4687298.40.082
Gradable, with mydriasis–––98.4–
Gradable, without mydriasis94.4––––
Excellent (score 3)45.8337262.30.031
Good (score 2)48.6357236.1–
Adequate (score 1)5.64721.6–
Ungradable (score 0)5.64721.6–
Both fields captured95.8697298.90.157
Single field only4.23721.1–
No artifacts76.4557287.20.048
Minor artifacts18.1137211.2–
Major artifacts5.64721.6–
Clear disc margins91.7667297.80.064
Clear vessel detail87.5637295.60.029
Clear macula93.1677298.30.074
Table 4. Subgroup analysis by disease severity, comorbidity and image quality (percent with bootstrap 95% CI).
SubgroupN (eyes)Sensitivity (%)Sensitivity 95% CI lowerSensitivity 95% CI upperSpecificity (%)Specificity 95% CI lowerSpecificity 95% CI upperPPV (%)PPV 95% CI lowerPPV 95% CI upperNPV (%)NPV 95% CI lowerNPV 95% CI upper
All referable DR3688.977.897.494.486.510094.184.899.389.579.396.1
Moderate NPDR1894.483.910094.486.510089.575.297.997.191.6100
Severe NPDR1283.362.697.794.486.510090.975.798.189.579.396.1
Proliferative DR683.352.59894.486.510083.352.59894.486.5100
DR only2290.98098.394.486.510095.286.59989.579.396.1
DR + DME1485.769.796.594.486.510092.379.198.489.579.396.1
Excellent quality3393.985.298.99791.510096.989.399.694.186.598.4
Good/adequate3585.773.894.991.482.397.190.98097.786.575.794.4
Table 5. Inter-grader agreement and reliability (72 eyes); Cohen's kappa with bootstrap 95% CI.
ComparisonAgreement (%)Agreeing eyesEyesCohen's kappaKappa 95% CI lowerKappa 95% CI upper
Grader 1 vs grader 2 (referable DR)95.869720.9170.8410.982
Grader 1 vs AI model (referable DR)91.766720.8330.7310.924
Grader 2 vs AI model (referable DR)90.365720.8060.6980.905
Grader 1 vs grader 2 (5-level severity)88.964720.8420.7520.926
Microaneurysms93.167720.8620.7730.941
Hemorrhages91.766720.8330.7360.924
Hard exudates95.869720.9170.8450.981
Cotton-wool spots97.270720.9430.8810.991
Neovascularization98.671720.9720.9270.998
Vitreous hemorrhage1007272111
Overall gradability97.270720.8310.6860.951
Quality score (4-point)88.964720.8220.7260.911

Access

How to cite

Pazo, E. E., Liu, X., Liu, S., Zhang, Q., Zhang, M., Zhang, Z., Moutari, S., Usama, M., Ren, X. (2026). Validation of the Eyerobo FC Portable Fundus Camera for Diabetic Retinopathy Screening Using Public Datasets and Deep Learning. Ophthalmology and Therapy. https://doi.org/10.1007/s40123-026-01353-w

Show BibTeX
@article{pazo2026validation,
  title = {{Validation of the Eyerobo FC Portable Fundus Camera for Diabetic Retinopathy Screening Using Public Datasets and Deep Learning}},
  author = {Pazo, Emmanuel Eric and Liu, Xiangying and Liu, Shoukuan and Zhang, Qinru and Zhang, Mingxuan and Zhang, Zhihui and Moutari, Salissou and Usama, Muhammad and Ren, Xinjun},
  journal = {Ophthalmology and Therapy},
  year = {2026},
  doi = {10.1007/s40123-026-01353-w},
  url = {https://doi.org/10.1007/s40123-026-01353-w},
}

Related work