 "title": "Validation of the Eyerobo FC Portable Fundus Camera for Diabetic Retinopathy Screening Using Public Datasets and Deep Learning",
 "url": "https://ericpazo.org/publications/2026-validation-of-the-eyerobo-fc-portable/",
 "doi": "10.1007/s40123-026-01353-w",
 "doi_url": "https://doi.org/10.1007/s40123-026-01353-w",
 "venue": "Ophthalmology and Therapy",
 "year": 2026,
 "authors": [
  "Emmanuel Eric Pazo",
  "Xiangying Liu",
  "Shoukuan Liu",
  "Qinru Zhang",
  "Mingxuan Zhang",
  "Zhihui Zhang",
  "Salissou Moutari",
  "Muhammad Usama",
  "Xinjun Ren"
 ],
 "topic": "Retina & vascular imaging",
 "design": "Prospective validation study",
 "sample": "104 eyes",
 "findings": [
  {
   "value": "92.3%",
   "label": "Sensitivity, portable camera",
   "proportion": 0.923
  },
  {
   "value": "94.2%",
   "label": "Specificity, portable camera",
   "proportion": 0.942
  },
  {
   "value": "0.977",
   "label": "AUC (desktop benchmark 0.952)",
   "proportion": 0.977
  }
 ],
 "conclusion": "A portable fundus camera with AI trained only on desktop images was noninferior to desktop systems for diabetic retinopathy screening.",
 "summary": "Prospective validation study of 104 eyes. Key results: Sensitivity, portable camera: 92.3%; Specificity, portable camera: 94.2%; AUC (desktop benchmark 0.952): 0.977. Conclusion: A portable fundus camera with AI trained only on desktop images was noninferior to desktop systems for diabetic retinopathy screening.",
 "images": {
  "png": "https://ericpazo.org/publications/2026-validation-of-the-eyerobo-fc-portable/visual-abstract.png",
  "svg": "https://ericpazo.org/publications/2026-validation-of-the-eyerobo-fc-portable/visual-abstract.svg",
  "width": 1200,
  "height": 630
 },
 "license": "https://creativecommons.org/licenses/by/4.0/",
 "source_note": "Finding values are quoted verbatim from the published abstract."
}