商用自适应光学扫描激光检眼镜(Mona II)测量健康成人视锥光感受器指标的可重复性

原文标题: Repeatability of cone photoreceptor metrics with a commercial adaptive optics scanning laser ophthalmoscopy (Mona II) in healthy adults

Jinyi Jia, Bei Du, Hua Rong, Biying Wang, Emmanuel Eric Pazo, Guihua Liu, Ruihua Wei

British Journal of Ophthalmology, 2026 · 期刊论文 · doi:10.1136/bjo-2025-329013 · 被引 1 次

视网膜与血管成像

图文摘要. 可重复性研究 · 126 名健康成人. 所有视锥指标的 ICC 范围: 0.967 to 0.994; 最高 ICC(视锥密度): 0.992. Mona II AO-SLO 在视网膜中央 4.2° 范围内测量视锥光感受器参数具有很高的可重复性。
图文摘要: SVG · PNG · JSON · 可在注明出处的前提下自由使用(CC BY 4.0)

要点速览

Mona II AO-SLO 在视网膜中央 4.2° 范围内测量视锥光感受器参数具有很高的可重复性。

研究设计
可重复性研究
样本
126 名健康成人
所有视锥指标的 ICC 范围
0.967 to 0.994
最高 ICC(视锥密度)
0.992

摘要

本文摘要以原文(英文)显示,因为该文章未采用允许翻译的开放许可协议。

PURPOSE: To evaluate the repeatability of the adaptive optics scanning laser ophthalmoscopy (AO-SLO) system (Mona II) in measuring cone photoreceptor parameters and compare repeatability differences in different retinal regions. METHODS: A total of 126 healthy adults aged between 18 and 36 years with cycloplegic spherical equivalent refraction ≤+0.50 D and >-6.00 D were enrolled. Retinal images were captured at the macular fovea centre (0°, 0°), the temporal, nasal, superior and inferior regions at 1.5° and 3.0° eccentricity. At each region, multiple 2.4°×2.4° images were sequentially captured with 0.9°×2.4° overlap between adjacent regions, covering the detection area within 4.2° centred on the fovea. Three images with clear cone photoreceptors and image quality scores of 0.8 or above were selected from each region for analysis. RESULTS: All parameters of the cone photoreceptors demonstrated excellent repeatability, with intraclass correlation coefficient (ICC) ranging from 0.967 to 0.994. The cone density exhibited good repeatability, with the fovea and the nasal region at 1.5° eccentricity showing the highest ICC (0.992). For the cone spacing, regularity and dispersion also demonstrated high repeatability across all regions (all ICC >0.90). CONCLUSIONS: The Mona II AO-SLO system showed high repeatability in measuring parameters of cone photoreceptors in healthy adults, with consistent performance across retinal area within a 4.2° range. TRIAL REGISTRATION NUMBER: ChiCTR2500114249.

关键词: cone photoreceptors; adaptive optics scanning laser ophthalmoscopy; repeatability; cone density; cone spacing; retinal eccentricity; healthy adults; fovea

获取全文

引用格式

Jia, J., Du, B., Rong, H., Wang, B., Pazo, E. E., Liu, G., Wei, R. (2026). Repeatability of cone photoreceptor metrics with a commercial adaptive optics scanning laser ophthalmoscopy (Mona II) in healthy adults. British Journal of Ophthalmology. https://doi.org/10.1136/bjo-2025-329013

显示 BibTeX
@article{jia2026repeatability,
  title = {{Repeatability of cone photoreceptor metrics with a commercial adaptive optics scanning laser ophthalmoscopy (Mona II) in healthy adults}},
  author = {Jia, Jinyi and Du, Bei and Rong, Hua and Wang, Biying and Pazo, Emmanuel Eric and Liu, Guihua and Wei, Ruihua},
  journal = {British Journal of Ophthalmology},
  year = {2026},
  doi = {10.1136/bjo-2025-329013},
  url = {https://doi.org/10.1136/bjo-2025-329013},
}

相关研究