Repetibilidad de las métricas de fotorreceptores de tipo cono con una oftalmoscopia láser de barrido con óptica adaptativa comercial (Mona II) en adultos sanos
Título original: Repeatability of cone photoreceptor metrics with a commercial adaptive optics scanning laser ophthalmoscopy (Mona II) in healthy adults
Jinyi Jia, Bei Du, Hua Rong, Biying Wang, Emmanuel Eric Pazo, Guihua Liu, Ruihua Wei
De un vistazo
El AO-SLO Mona II mide los parámetros de los fotorreceptores de tipo cono con alta repetibilidad en los 4.2° centrales de la retina.
- Diseño
- Estudio de repetibilidad
- Muestra
- 126 adultos sanos
- Rango de ICC, todas las métricas de conos
- 0.967 to 0.994
- ICC más alto (densidad de conos)
- 0.992
Resumen
Este resumen se muestra en su idioma original, el inglés, porque el artículo no tiene una licencia abierta que permita traducciones.
PURPOSE: To evaluate the repeatability of the adaptive optics scanning laser ophthalmoscopy (AO-SLO) system (Mona II) in measuring cone photoreceptor parameters and compare repeatability differences in different retinal regions. METHODS: A total of 126 healthy adults aged between 18 and 36 years with cycloplegic spherical equivalent refraction ≤+0.50 D and >-6.00 D were enrolled. Retinal images were captured at the macular fovea centre (0°, 0°), the temporal, nasal, superior and inferior regions at 1.5° and 3.0° eccentricity. At each region, multiple 2.4°×2.4° images were sequentially captured with 0.9°×2.4° overlap between adjacent regions, covering the detection area within 4.2° centred on the fovea. Three images with clear cone photoreceptors and image quality scores of 0.8 or above were selected from each region for analysis. RESULTS: All parameters of the cone photoreceptors demonstrated excellent repeatability, with intraclass correlation coefficient (ICC) ranging from 0.967 to 0.994. The cone density exhibited good repeatability, with the fovea and the nasal region at 1.5° eccentricity showing the highest ICC (0.992). For the cone spacing, regularity and dispersion also demonstrated high repeatability across all regions (all ICC >0.90). CONCLUSIONS: The Mona II AO-SLO system showed high repeatability in measuring parameters of cone photoreceptors in healthy adults, with consistent performance across retinal area within a 4.2° range. TRIAL REGISTRATION NUMBER: ChiCTR2500114249.
Palabras clave: cone photoreceptors; adaptive optics scanning laser ophthalmoscopy; repeatability; cone density; cone spacing; retinal eccentricity; healthy adults; fovea
Acceso
Cómo citar
Mostrar BibTeX
@article{jia2026repeatability,
title = {{Repeatability of cone photoreceptor metrics with a commercial adaptive optics scanning laser ophthalmoscopy (Mona II) in healthy adults}},
author = {Jia, Jinyi and Du, Bei and Rong, Hua and Wang, Biying and Pazo, Emmanuel Eric and Liu, Guihua and Wei, Ruihua},
journal = {British Journal of Ophthalmology},
year = {2026},
doi = {10.1136/bjo-2025-329013},
url = {https://doi.org/10.1136/bjo-2025-329013},
}
Trabajos relacionados
- Asociación entre la presión de perfusión ocular media y la retinopatía diabética proliferativa en una población del norte de China (2026)
- Asociación de los conocimientos y actitudes sobre el tratamiento anti-VEGF con la interrupción autodeclarada del tratamiento: un estudio de clasificación con separación temporal (2026)
- Medición automatizada basada en aprendizaje profundo del volumen de retinosquisis y desprendimiento en la miopía patológica con contracción escleral posterior (2026)
- Desarrollo de un modelo predictivo basado en datos de citocinas del humor acuoso para la respuesta al tratamiento anti-VEGF en el edema macular diabético (2026)
- Cambios retinianos y coroideos en pacientes diabéticos antes del diagnóstico definitivo de retinopatía diabética: un estudio prospectivo de dos años (2026)