2WIN‐S 便携式验光仪与睫状肌麻痹检影验光在学龄儿童屈光测量中的比较:一项横断面研究
原文标题: Comparing School‐Aged Refraction Measurements Using the 2WIN‐S Portable Refractor in Relation to Cycloplegic Retinoscopy: A Cross‐Sectional Study
Ziming Liu, Emmanuel Eric Pazo, Hong Ye, Cui Yu, Ling Xu, Wei He
要点速览
作为筛查工具,2WIN-S 摄影筛查仪与儿童睫状肌麻痹检影验光的一致性良好。
- 研究设计
- 横断面一致性研究
- 样本
- 97 名儿童(194 只眼)
- ICC(等效球镜)
- 0.900
- 位于一致性界限内
- 95%
摘要
目的:评估 2WIN-S 摄影筛查屈光测量的重复性,以及其与金标准睫状肌麻痹检影验光的一致性。设计:单中心横断面研究。方法:使用检影镜和 2WIN-S 评估 97 名儿童双眼共 194 只眼的球镜、柱镜、轴位和等效球镜。一周后,由另一名操作者重复 2WIN-S 测量。主要结局指标为 2WIN-S 等效球镜、J0 和 J45 读数的重复性和一致性。测量重复性采用受试者内标准差(2.77 Sw)和组内相关系数(ICC)评估;设备间一致性采用 95% 一致性界限评估;睫状肌麻痹检影验光与非睫状肌麻痹 2WIN-S 测量之间的一致程度采用 Bland-Altman 分析评估。结果:value >0.86)。SE、J0 和 J45 的 ICC 分别为 0.900、0.666 和 0.639;SE、J0 和 J45 的 Sw 分别为 0.61D、0.30D 和 0.31D;睫状肌麻痹检影验光与 2WIN-S 的 SE Bland-Altman 分析显示,184/194(95%)位于 95% 置信区间内,均值为 0.46。J0 为 184/194(95%),均值为 -0.04。J45 为 181/194(93%),均值为 -0.15。结论:在儿童和青少年中,2WIN-S 的客观屈光测量具有良好的信度,并与金标准检影验光高度一致。尽管观察到良好的一致性,但须考虑到它是一种筛查工具。
译自英文原文摘要(许可协议:CC BY)。
显示英文原文摘要
PURPOSE: To assess the repeatability and agreement of refractive measurements using 2WIN-S photoscreening with the gold-standard cycloplegic retinoscope refraction. DESIGN: Single centre, cross-sectional study. METHODS: Spherical, cylindrical, axis, and spherical equivalent of 194 bilateral eyes of 97 children were assessed using a retinoscope and 2WIN-S. One week later, another operator repeated the 2WIN-S measurements. The primary outcome measures were to assess the repeatability and agreement between spherical equivalent, J0, and J45 readings of 2WIN-S. The repeatability of measurements was assessed by the within-subject standard deviation (2.77 Sw) and intraclass correlation coefficient (ICC). The agreement between devices was assessed using 95% limits of agreement. The extent of the agreement between cycloplegic retinoscopy and noncycloplegic 2WIN-S measurements was assessed using Bland-Altman analysis. RESULTS: value >0.86). ICC for SE, J0, and J45 was 0.900, 0.666, and 0.639, respectively; Sw for SE, J0, and J45 was 0.61D, 0.30D, and 0.31D, respectively; Bland-Altman analysis of retinoscopy with cycloplegia and 2WIN-S for SE was 184/194 (95%) in 95% confidence interval, and the mean value was 0.46. J0 was 184/194 (95%), and the mean value is -0.04. J45 was 181/194 (93%), and the mean value is -0.15. CONCLUSION: The objective refractive measurement of 2WIN-S had good reliability and high agreement with the gold-standard retinoscopy refraction in children and adolescents. While consistency was observed, it is essential to take into consideration that it is a screening tool.
关键词: cycloplegic retinoscopy; refractive error measurement; children and adolescents; repeatability; spherical equivalent; astigmatism axis; non-cycloplegic refraction; school vision screening
获取全文
引用格式
显示 BibTeX
@article{liu2021comparing,
title = {{Comparing School‐Aged Refraction Measurements Using the 2WIN‐S Portable Refractor in Relation to Cycloplegic Retinoscopy: A Cross‐Sectional Study}},
author = {Liu, Ziming and Pazo, Emmanuel Eric and Ye, Hong and Yu, Cui and Xu, Ling and He, Wei},
journal = {Journal of Ophthalmology},
year = {2021},
doi = {10.1155/2021/6612476},
url = {https://doi.org/10.1155/2021/6612476},
}