2WIN‐S 便携式验光仪与睫状肌麻痹检影验光在学龄儿童屈光测量中的比较:一项横断面研究

原文标题: Comparing School‐Aged Refraction Measurements Using the 2WIN‐S Portable Refractor in Relation to Cycloplegic Retinoscopy: A Cross‐Sectional Study

Ziming Liu, Emmanuel Eric Pazo, Hong Ye, Cui Yu, Ling Xu, Wei He

Journal of Ophthalmology, 2021 · 期刊论文 · doi:10.1155/2021/6612476 · 被引 15 次 · 开放获取(CC BY)

近视与屈光不正视网膜与血管成像

图文摘要. 横断面一致性研究 · 97 名儿童(194 只眼). ICC(等效球镜): 0.900; 位于一致性界限内: 95%. 作为筛查工具,2WIN-S 摄影筛查仪与儿童睫状肌麻痹检影验光的一致性良好。
图文摘要: SVG · PNG · JSON · 可在注明出处的前提下自由使用(CC BY 4.0)

要点速览

作为筛查工具,2WIN-S 摄影筛查仪与儿童睫状肌麻痹检影验光的一致性良好。

研究设计
横断面一致性研究
样本
97 名儿童(194 只眼)
ICC(等效球镜)
0.900
位于一致性界限内
95%

摘要

目的:评估 2WIN-S 摄影筛查屈光测量的重复性,以及其与金标准睫状肌麻痹检影验光的一致性。设计:单中心横断面研究。方法:使用检影镜和 2WIN-S 评估 97 名儿童双眼共 194 只眼的球镜、柱镜、轴位和等效球镜。一周后,由另一名操作者重复 2WIN-S 测量。主要结局指标为 2WIN-S 等效球镜、J0 和 J45 读数的重复性和一致性。测量重复性采用受试者内标准差(2.77 Sw)和组内相关系数(ICC)评估;设备间一致性采用 95% 一致性界限评估;睫状肌麻痹检影验光与非睫状肌麻痹 2WIN-S 测量之间的一致程度采用 Bland-Altman 分析评估。结果:value >0.86)。SE、J0 和 J45 的 ICC 分别为 0.900、0.666 和 0.639;SE、J0 和 J45 的 Sw 分别为 0.61D、0.30D 和 0.31D;睫状肌麻痹检影验光与 2WIN-S 的 SE Bland-Altman 分析显示,184/194(95%)位于 95% 置信区间内,均值为 0.46。J0 为 184/194(95%),均值为 -0.04。J45 为 181/194(93%),均值为 -0.15。结论:在儿童和青少年中,2WIN-S 的客观屈光测量具有良好的信度,并与金标准检影验光高度一致。尽管观察到良好的一致性,但须考虑到它是一种筛查工具。

译自英文原文摘要(许可协议:CC BY)。

显示英文原文摘要

PURPOSE: To assess the repeatability and agreement of refractive measurements using 2WIN-S photoscreening with the gold-standard cycloplegic retinoscope refraction. DESIGN: Single centre, cross-sectional study. METHODS: Spherical, cylindrical, axis, and spherical equivalent of 194 bilateral eyes of 97 children were assessed using a retinoscope and 2WIN-S. One week later, another operator repeated the 2WIN-S measurements. The primary outcome measures were to assess the repeatability and agreement between spherical equivalent, J0, and J45 readings of 2WIN-S. The repeatability of measurements was assessed by the within-subject standard deviation (2.77 Sw) and intraclass correlation coefficient (ICC). The agreement between devices was assessed using 95% limits of agreement. The extent of the agreement between cycloplegic retinoscopy and noncycloplegic 2WIN-S measurements was assessed using Bland-Altman analysis. RESULTS: value >0.86). ICC for SE, J0, and J45 was 0.900, 0.666, and 0.639, respectively; Sw for SE, J0, and J45 was 0.61D, 0.30D, and 0.31D, respectively; Bland-Altman analysis of retinoscopy with cycloplegia and 2WIN-S for SE was 184/194 (95%) in 95% confidence interval, and the mean value was 0.46. J0 was 184/194 (95%), and the mean value is -0.04. J45 was 181/194 (93%), and the mean value is -0.15. CONCLUSION: The objective refractive measurement of 2WIN-S had good reliability and high agreement with the gold-standard retinoscopy refraction in children and adolescents. While consistency was observed, it is essential to take into consideration that it is a screening tool.

关键词: cycloplegic retinoscopy; refractive error measurement; children and adolescents; repeatability; spherical equivalent; astigmatism axis; non-cycloplegic refraction; school vision screening

获取全文

引用格式

Liu, Z., Pazo, E. E., Ye, H., Yu, C., Xu, L., He, W. (2021). Comparing School‐Aged Refraction Measurements Using the 2WIN‐S Portable Refractor in Relation to Cycloplegic Retinoscopy: A Cross‐Sectional Study. Journal of Ophthalmology. https://doi.org/10.1155/2021/6612476

显示 BibTeX
@article{liu2021comparing,
  title = {{Comparing School‐Aged Refraction Measurements Using the 2WIN‐S Portable Refractor in Relation to Cycloplegic Retinoscopy: A Cross‐Sectional Study}},
  author = {Liu, Ziming and Pazo, Emmanuel Eric and Ye, Hong and Yu, Cui and Xu, Ling and He, Wei},
  journal = {Journal of Ophthalmology},
  year = {2021},
  doi = {10.1155/2021/6612476},
  url = {https://doi.org/10.1155/2021/6612476},
}