 "title": "Automatic measurement and evaluation of anterior segment anatomical structures via UBM images using a deep learning-based approach",
 "url": "https://ericpazo.org/publications/2026-automatic-measurement-and-evaluation-of-anterior/",
 "doi": "10.1007/s00417-025-07070-7",
 "doi_url": "https://doi.org/10.1007/s00417-025-07070-7",
 "venue": "Graefe's Archive for Clinical and Experimental Ophthalmology",
 "year": 2026,
 "authors": [
  "S Chen",
  "X Chu",
  "EE Pazo",
  "Y Huang",
  "H Liu",
  "C Zhang",
  "R Yang",
  "S Zhao"
 ],
 "topic": "AI & deep learning",
 "design": "Cross-sectional retrospective study",
 "sample": "638 eyes",
 "findings": [
  {
   "value": ">0.978",
   "label": "ICC vs reference standard",
   "proportion": 0.978
  },
  {
   "value": "<1.7%",
   "label": "Mean relative error",
   "proportion": 0.017
  },
  {
   "value": "1,480",
   "label": "UBM images analyzed",
   "proportion": null
  }
 ],
 "conclusion": "A YOLOv8-pose model measures anterior segment parameters on UBM images as accurately as experienced clinicians, aiding ICL sizing.",
 "summary": "Cross-sectional retrospective study of 638 eyes. Key results: ICC vs reference standard: >0.978; Mean relative error: <1.7%; UBM images analyzed: 1,480. Conclusion: A YOLOv8-pose model measures anterior segment parameters on UBM images as accurately as experienced clinicians, aiding ICL sizing.",
 "images": {
  "png": "https://ericpazo.org/publications/2026-automatic-measurement-and-evaluation-of-anterior/visual-abstract.png",
  "svg": "https://ericpazo.org/publications/2026-automatic-measurement-and-evaluation-of-anterior/visual-abstract.svg",
  "width": 1200,
  "height": 630
 },
 "license": "https://creativecommons.org/licenses/by/4.0/",
 "source_note": "Finding values are quoted verbatim from the published abstract."
}